Advertisement

5

5.2 Transporting GasLab Q2

PROHIBITION INTERDICTION
Remove of transport guards and other safety devices, for example,
l
for gas connections, before the device has been installed at its
place of use and if dirt can still penetrate it.
Retirer les protections de transport et autres dispositifs de sécurité,
par exemple pour les raccords de gaz, avant que l'appareil n'ait été in-
stallé sur son lieu d'utilisation et si la saleté peut encore y pénétrer.
Prerequisite(s)
Transport protection guards and protective devices, e.g. for gas connections,
l
are attached to the instrument.
Steps
1. Transport the device to its place of use, avoiding extreme temperature fluc-
tuations and the formation of condensation.
2. Check the device for condensation, if no condensation has occurred during
transport the device can be used immediately. Otherwise, continue to the next
step.
3. Slowly bring the device to room temperature.
4. Open the device
12 hours. If necessary, contact Honeywell.
GasLab Q2
GasLab Q2 after delivery and at the place of use
5.5.1 Opening GasLab Q2
➪ 
User Manual (N )
(p. 70) and let it dry for at least