Most applications require only 2 cable glands, one for the data transfer cable and
one for the power supply (24 V DC). The remaining holes provide access for further
signal cables if required. Select a suitable type of cable gland depending on the loc-
ation and requirements, contact Honeywell if necessary.
In order to maintain the specified explosion protection, the screw-in parts must be
installed and cast in accordance with the gland manufacturer's specifications. Com-
pliance with the IP rating may require the use of sealing rings.
Openings which are not required must be fitted with appropriate dummy plugs.
DANGER due to incorrect cable entry devices.
DANGER en raison de dispositifs d'entrée de câbles incorrects.
The cable entries must meet or exceed the IP rating (IP 64). Depending
on the type, additional sealing rings and washers must be used.
Les entrées de câbles doivent être conformes ou supérieures à l'indice IP
(IP 64). Selon le type, des bagues d'étanchéité et des rondelles sup-
plémentaires doivent être utilisées.
GasLab Q2
GasLab Q2 after delivery and at the place of use
Fig. 5-5: Position of the cable glands
User Manual (N )
79 of 315