10.4.3 Remove of the battery
For long time storing or disposal the battery has to be removed. If the battery
charge was too low according to the display, or a new board is delivered without bat-
tery, a replacement lithium battery must be inserted.
Prerequisite(s)
Opening of device is permitted by plant operator.
l
Ambient conditions according to ISO / IEC61010-1
l
(temperatures / humidity as in living and office spaces).
The parameterization of the device was read out with enSuite,
l
meterization
(p. 182)
The following tools and materials are available:
Service standard tools
l
Gas leak detector for flammable gases, calibrated with status or value display
l
ESD wrist strap or similar device to prevent electrostatic discharge
l
Steps
Preparations for opening and opening the device:
1. Dee-energize the device (switch off power).
2. Make sure that there is no voltage, e.g. with a multimeter.
3. Turn off gas supply, block the device from all gases.
4. Monitor the environment for the presence of an explosive atmosphere, e.g., with
a gas leak detector, continuously during the entire process.
5. Ensure there is no explosive atmosphere. Only then open the device.
5.5.1 Opening GasLab Q2
➪
DANGER EXPLOSIVE ATMOSPHERE / HOT SURFACES
Damage to health and property (explosion) possible
Stop work immediately if an explosive atmosphere occurs!
The sensor block is heated up to 70 °C!
Let it cool down, be careful not to burn yourself.
DANGER ATMOSPHÈRE EXPLOSIVE / SURFACES CHAUDES
Possibilité d'atteinte à la santé et aux biens (explosion)
Arrêtez immédiatement le travail en cas d'atmosphère explosive !
Le bloc capteur est chauffé jusqu' à 70°C ≙158°F !
Laissez-le refroidir, en veillant à ne pas vous brûler.