GasLab Q2 must be installed so that the main breather element points down-
wards and the cable connections point upwards.
GasLab Q2 does not require a special mounting plate since the measurement
device contains all the parts required for the measuring task in its explosion-pro-
tected housing. It may be installed on a flat vertical surface (wall/measuring board).
Use the 8 mm ≙0.31 inch fastening holes provided in the housing for this purpose.
Use size M8 (minimum M6) fastening material or equivalent in all four fastening
holes to secure the device. Suitable complements (in other words, nuts or anchors)
depend on the material and the type of installation surface.
The material to which the unit is attached must be able to withstand a combined
vertical load of 1000 N at the four attachment points, equivalent to a weight of ap-
proximately 100 kg.
DANGER Working in hazardous areas
DANGER Travailler dans des zones dangereuses
Risk of damage to health, environment and plant
Risque d'atteinte à la santé, à l'environnement et à l'installation
Obey the explosion-protection rules at all times when working with tools
or on items in hazardous zones.
Respectez à tout moment les règles de protection contre les explosions
lorsque vous travaillez avec des outils ou sur des objets dans des zones
dangereuses.