4. Select the Transfer parameterization to device action (for example in the con-
text menu).
5. Log in using the login dialog with your user name and password which appears.
The parameter set will be transferred.
There are typically a number of parameters which are protected by the security
switch (SSW) (e.g. in order to meet the requirements of an approval).
If the new device parameterization is likely to bring about changes to such para-
meters, but the security switch is closed, then the message Device could not be
parameterized appears following the transfer. The device does not restart; the old
parameterization is still used.
If the parameters are not protected by the SSW or if this is open, the changes will be
transferred, and the device will restart.
Fig. 8-41: Dialog parameterization transferred successfully
The restart will disconnect the device and the user will be logged out automati-
cally.
The device will now operate using the transferred settings after you have re-enabled
the required operating mode. These new settings will remain in use even after the
device is switched off and on again.
GasLab Q2
Configuration and analysis software enSuite
Fig. 8-40: Dialog transfer parameterization
User Manual (N )
231 of 315