Operating mode: Operational calibration
(manual)
The last valid gas property values are held throughout the calibration
time using measurement outputs. The parameters from the last
successful calibration remain in use until a new successful cal-
ibration has taken place. Errors will be displayed on the device and
documented in the "Error List".
Prerequisite(s)
The device is in automatic normal operation (Analysis).
l
Sensors are in the measuring range. (no alarm / no warning)
l
Calibration gas cylinder is properly connected and ready for use
l
You are logged in with the necessary rights.
l
Steps
1. .Go to: Home
2. After a change in the parameterization or with a new calibration gas, click on
"Gas components" in the "Operational calibration" line.
Check or correct the calibration gas mixture using the manufacturer's analysis
certificate. See
➪  
3. Activate "Start" in the "Operational Calibration" line
(only possible if all requirements are met)
4. Click on "Confirm". The procedure now runs automatically, the warning "Oper-
ational Calibration" is active during this time. By selecting "Cancel"you can end
the
GasLab Q2
-> Control ->
Calibration
7.2.6 Entries and changes using the operation panel
calibration
User Manual (N )
GasLab Q2 operation panel and operation
Login
(p. 135)
➪ 
at
any
.
(p. 112)
time.