9. You will be requested to restore the original gas connection.
(The message "Prepare process gas" will appear depending on the actual input
used. See figure)
10. Restore the original connection correctly. Obey all the safety information for
working on gas connections. Conduct a tightness test. See section
Check the tightness of the installation
11. Once the gas has been connected, acknowledge the security query from step 9
(see figure above) with Confirm. The device will revert to analysis mode, and the
verification is ended.
If you log out during the task, you will no longer be able to access the
verification function using the operation panel. The last confirmed step
will still be performed. Log in again to continue.
GasLab Q2
GasLab Q2 operation panel and operation
(p. 264).