10.4.1 Connect/replace calibration-, verification- gas
cylinders
For proper operation of the instrument, regular automatic calibration/adjustment
with a calibration gas is necessary. Refer to section
measuring device
(p. 65)
As soon as the outlet pressure of a gas cylinder can no longer be kept stable, it
must be replaced. Depending on the regulator, this may occur at a pressure around
1.0 MPa (g). If the pressure is below 2 MPa, you should have a replacement cylinder
ready. therefore monitor the cylinder pressure with e.g. a pressure gauge or contact
manometer.
The recognized rules of technology for handling gases must be observed for all
work as described. Surfaces that come into contact with gas must be free of grease,
oil, solvents and other contaminants. Take steps to prevent dirt and/or moisture
from reaching the unit during replacement.
DANGER
Incorrect handling of gas cylinders
Manipulation incorrecte des bouteilles de gaz
Serious damage to health and considerable damage to property can
occur (explosion) De graves dommages à la santé et des dommages
matériels considérables peuvent survenir (explosion).
Complete the following steps in accordance with national regulations on
the basis of the information in this document.Suivez impérativement les
étapes suivantes conformément aux normes locales en respectant les in-
structions de ce manuel.
Remove connected gas cylinder
1. Close the main valve on the gas cylinder.
2. Depressurize the connected gas pipeline.
3. Remove everything from the gas cylinder connector.
4. Close the connection thread on the valve with the screw cover (nut).
5. Place the protective cap onto the gas cylinder and screw it tightly.
6. Release the anti-tipping device and remove the gas cylinder.
7. Protect the connection thread for the remaining installation if a new gas cylinder is not to be con-
nected immediately.