Honeywell enCore Series Manual page 168

Gas quality analyzer
Hide thumbs Also See for enCore Series:
Table of Contents

Advertisement

7
Steps
1. Activate on the device or in the remote operation panel
(Home-> Q2 Control)
2. Activate "References". Check whether the test gas mixture entered corres-
ponds to the one used, if not, adjust it.
operation panel
3. Go back to
4. The analysis (current measurement) will no longer be updated or will be
stopped. The device waits for a confirmation entry.
5. Change the gas at the parameterized input (process gas (PG) or calibration gas
input (CAL)
8.4.1 Setting or changing automatic measurements
➪  
connect the cylinder with the test gas to the inlet via a cylinder pressure reducer.
Note
2 Safety information (EN-US-FRA)
➪  
calibration-, verification- gas cylinders
6. Open the test gas cylinder and set the pressure between 0.15 MPag ≙1.5 barg
and 0.3 MPag ≙3 barg.
GasLab Q2
(p. 112)
and activate "Start"
User Manual (N )
GasLab Q2 operation panel and operation
7.2.6 Entries and changes using the
➪  
(p. 16)
10.4.1 Connect/replace
➪  
(p. 276)
(p. 236) and
168 of 315