Disconnecting device and cables

//Cette page est répétée en français. // This page is repeated in French. //
The device does not have its own off switch and must be operated by the customer
via a disconnecting device and current limiting device (or combination of both) to
comply with ISO / IEC60079-14 and ISO / IEC61010-1.
An example of such a disconnecting device, which is not included in the scope of
supply, is a C 4A circuit breaker, which must be installed near the Q2 for reasons of
explosion protection and electrical safety.
The disconnect device must meet the requirements of the installation site, must be
capable of safely disconnecting at least 48 V DC, and must disconnect all power
supply lines leading into the hazardous area. It thus contributes to the formation of
the power-limited circuit.
DANGER / WARNING / CAUTION
Use of wrong equipment Installation errors
Health hazard Explosion hazard
The installation personnel must ensure that the wiring is protected from
overloads and short-circuits and has the required over voltage and over
current protection.
The supply and communication equipment (cables not supplied) must
be screened and comply with local installation regulations.
The cable shall be suitable for + 80°C ≙176°F.
PROHIBITION
Operation without disconnection device
l
Replacement of internal fuses without authorization
l
Without a disconnect device, the unit no longer complies with ISO /
IEC60079- 14 and ISO / IEC61010- 1. Replacement of internal fuses
may only be performed by Honeywell authorized personnel.
GasLab Q2
Safety information (EN-US-FRA)
User Manual (N )
35 of 315