Lighting area adjustment [G-LS]

5
It is possible to adjust the lighting area by moving the lamp socket back
and forth. Adjust the lighting area to allow easy viewing.

Attaching filters [C-DSLS]

6
Rotate and remove the filter cap at the end of
the lamp house and insert a filter. Use filters
that have a diameter of 33 mm and a width of
1.5 ~ 2 mm.
2
P-ICI2 Coaxial Episcopic Illuminator
Main body
Eyepiece tube
clamp screw
Luminous section
Incident section
Magnification of the P-ICI2 Coaxial
Episcopic Illuminator is 1.5X. Total
magnification is obtained by multiplying 1.5
by the total magnification of the basic set of
the microscope.
Note) Usable magnification for the P-ICI2
Coaxial Episcopic Illuminator is
restricted. Use the magnification
within the range shown in the right
Light guide mount
Light guide clamp screw
Light guide
Illuminator
Filter
ø33 mm
Filter cap
Fiber illuminator
Fiber illuminator (C-FI115 for
100/120V region and C-FI230
for 230V region) available on
order is necessary for the
light source.
Read also the Instruction
Manual of the Fiber
Illuminator C-FI115/C-FI230.
Zooming
Objective
magnification
Acro 0.5X
3X or higher
Plan 1X
1.5X or higher
Plan 1X ERGO
1.5X or higher
Plan Apo 0.5X
4X or higher
Plan Apo 1X
2X or higher
ED Plan 1.5X
2X or higher
ED Plan 2X
Not usable
HR Plan Apo 0.5X
2.5X or higher
HR Plan Apo 1X
2.5X or higher
HR Plan Apo 1.6X
Not usable