Verification - Functional Checks - GE Vivid 7 Service Manual

Hide thumbs Also See for Vivid 7:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FC091194, R
IRECTION
EVISION
8-8-12-3
Verify TCP/IP Settings - Software Version v3.x
1.) From the top of the connectivity tab, select the Tcpip tab.
2.) Verify that all settings on the TCPIP tab matches those recorded in
Tcpip screen (v3.x)" on page
3.) Select Save settings.
4.) Select Ok to continue.
8-8-12-4
DC Offset Calibration
1.) Disconnect all probes.
2.) From the upper section of the window, select Test.
3.) Then select Go to System Test.
4.) From the Module selection block, select DC Offs. Cal.
5.) Reboot the System when the DC Offset Calibration has been completed.
8-8-12-5
Verify System Software Version
1.) Press CONFIG (F2) and log on as adm, see
2.) Select about tab.
3.) Verify that the SW versions are the same as what were recorded in
and Hardware versions (v3.x)" on page
8-8-13

Verification - Functional Checks

Please perform the tests specified in
11
Figure 8-48 TCP/IP Example Settings
8-44.
Chapter 8 - Replacement Procedures
4-2-4 "Log On to the System as 'ADM'" on page
8-46.
Section 8-20 "Verification - Functional Checks" on page
V
7 S
M
IVID
ERVICE
Table 8-11 "Record settings from
Table 8-13 "Record Software
ANUAL
4-12.
8-105.
8 - 53