Software Loading - Software Version v7.x/v6.x/v5.x/v4.x
8-7-12-1
Overview
This software loading procedure covers:
Re-installation of BT'08 software (System Software v3.0.x and Application Software v7.x) on a
Vivid 7 unit with BEP4.3
Re-installation of BT'08 software (System Software v2.1.1 and Application Software v7.x) on a
Vivid 7 unit with BEP3.x or BEP4.2
Re-installation of BT'06 software (System Software v2.x.x and Application Software v6.x) on a
Vivid 7 unit.
Re-installation of BT'05 software (System Software v2.x.x and Application Software v5.x) on a
Vivid 7 unit.
Re-installation of BT'04 software (System Software v2.x.x and Application Software v4.x) on a
Vivid 7 unit.
Refer to
where you recorded the unit's software versions.
The software for Vivid 7 is divided on two CDs:
one CD with the System Software
one CD with the Application Software
For a software CD overview, see
8-7-12-2
Introduction to Software Reloading
Starting with System Software version v3.0.0 (and Application Software v7.0.0), the compressed
System and Application software has been stored on the Vivid 7's hard disk drive, making a
software reinstallation easier and faster than with previous software.
-
-
For Vivid 7 with System Software version v2.x.x, the software CDs must be used if the software
needs to be reinstalled.
-
-
8 - 24
Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x
11
Table 8-8 "Record Software versions - Software Version v7.x/v6.x/v5.x/v4.x" on page 8-23
To reinstall the Application Software, go to:
8-7-12-4 "Loading Application Software v7.x on Vivid 7 with BEP4.3" on page
To reinstall both the System Software and the Application Software, go to:
8-7-12-3 "Loading System Software v3.0.x on Vivid 7 with BEP4.3" on page
To reinstall the Application Software, go to:
8-7-12-6 "Application SW Loading - Software Version v7.x/v6.x/v5.x/v4.x" on page 8-32
To reinstall both the System Software and the Application Software, go to:
8-7-12-5 "Vivid 7 System Software Loading - Application Software Version v7.x/v6.x/v5.x/
v4.x" on page 8-29