GE H
EALTHCARE
D
FC091194, R
IRECTION
EVISION
Table 2-4
GE VINGMED
PART NUMBER
FC000210
Vivid 7 (BT '02)
FC000200
Vivid 7 (BT '02)
FC000190
Vivid 7 PRO (BT '02)
FC000180
Vivid 7 PRO (BT '02)
FC000060
Vivid 7 (BT '01)
FC000030
Vivid 7 (BT '01)
2-2-2-3
Inrush Current
Inrush current is not a factor to consider due to the inrush current limiting properties of the power
supplies.
Maximum inrush current at startup:
230 VAC: 12 A
120 VAC: 7 A
2-2-2-4
Site Circuit Breaker
POWER OUTAGE MAY OCCUR. The Vivid 7 requires a dedicated single branch circuit. To avoid
CAUTION
circuit overload and possible loss of critical care equipment, make sure you DO NOT have any
other equipment operating on the same circuit.
It is recommended that the branch circuit breaker for the unit be readily accessible.
2-2-2-5
Site Power Outlets
A dedicated AC power outlet must be within reach of the unit without extension cords. Other outlets
adequate for the external peripherals, medical and test equipment needed to support this unit must also
be present within 1 m (3.2 ft.) of the unit. Electrical installation must meet all current local, state, and
national electrical codes.
2-2-2-6
Unit Power Plug
If the unit arrives without a power plug, or with the wrong plug, you must contact your GE dealer or the
installation engineer must supply what is locally required.
2-2-2-7
Power Stability Requirements
Voltage drop-out
Max 10 ms.
Power Transients
(All applications)
Less than 25% of nominal peak voltage for less than 1 millisecond for any type of transient, including
line frequency, synchronous, asynchronous, or aperiodic transients.
2 - 6
11
Electrical Specifications for Vivid 7 (cont'd) sheet 3 of 3
DESCRIPTION
Section 2-2 - General Console Requirements
VOLTAGE
TOLERANCES
100-120 VAC
±10%
230 VAC
±10%
100-120 VAC
±10%
230 VAC
±10%
100-120 VAC
±10%
230 VAC
±10%
V
7 S
M
IVID
ERVICE
ANUAL
CURRENT
FREQUENCY
10 A
50-60 Hz
5 A
50-60 Hz
10 A
50-60 Hz
5 A
50-60 Hz
10 A
50-60 Hz
5 A
50-60 Hz