Create a New Dataflow with Worklist and DICOM Storage (for CA-1000 sites)
This dataflow is intended for Vivid 7 scanners in a DICOM network, where the data should not be stored
locally on the scanner. A site with the integrated EchoPAC PC and CA-1000 is an example of this.
To create a new dataflow with Worklist and DICOM Storage, do the following steps:
1.) Add a new dataflow in Config screen, Connectivity, Dataflow tab. Enter a name, e.g. Worklist -
DICOM Storage.
2.) Select OK.
3.) Add the Worklist input. Configure (at least) IP address, AE title and Port Number.
4.) Select OK.
5.) Add the Dicom Storage output.
6.) Configure (at least) IP address, AE title and Port Number. If the site has integrated EchoPAC PC
and CA-1000, enable Allow SR and Allow SR Private Data.
7.) Select OK.
8.) If this dataflow should be the default at start-up, enable Default for the dataflow.
11