Preparing For Setup; Verifying Customer Order; Physical Inspection; Component Inspection - GE Venue Fit Technical Publication

Table of Contents

Advertisement

P R E L I M I N A R Y
DIRECTION 5854713-100, REVISION 2
Section 3-4

Preparing for Setup

3-4-1

Verifying Customer Order

Compare items received by the customer to that which is listed on the delivery order. Report any items
that are missing, back ordered, or damaged.
3-4-2

Physical Inspection

Verify that the Venue Fit™ arrived intact (visual inspection).
If the Venue Fit™ has been damaged, please refer to
the beginning of this manual.
3-4-3

Component Inspection

After verifying that all the required parts are included in the shipping crate, inspect the system components
using the checklist supplied below. In addition, ensure that all the labels described in
Introduction
label into the system installation details card, as described in Paperwork After Setup on page 3-53.
3-4-3-1

Damage Inspection Checklist

Visually inspect the contents of the shipping carton for damage. If any parts are damaged or missing,
contact an authorized GE Service Representative.
A Damage Inspection Checklist for the Venue Fit™ portable ultrasound scanner is provided in
11
below.
Table 3-11
Step
are present, accurate and in good condition, and enter the serial number printed on the main
Damage Inspection Checklist - Venue Fit™ Systems
Item
Venue Fit™
Verify that the system is switched OFF and unplugged. Clean the system. Check
1
System
that there are no dents or scratches and that no internal parts are exposed.
Check all probes for wear and tear on the lens, cable, and connector. Look for bent
or damaged pins on the connector and in the connector socket on the unit. Verify
2
Probes
that the EMI fingers around the probe connector socket housing are intact. Check
the probe locking mechanism and probe switch.
Check the rear panel connectors for bent pins, loose connections and loose or
3
Rear Panel
missing hardware. Screw all the cable connectors tightly to the connector sockets
on the panel. Verify that the labeling is in good condition.
Check the power cord for cuts, loose hardware, tire marks, exposed insulation, or
4
Power Cord/s
any deterioration. Verify continuity.
5
Venue Fit™ Cart
Check that there are no dents or scratches and that no internal parts are exposed.
Check and clean the peripherals in accordance with the manufacturer's directions.
6
Peripherals
To prevent EMI or system overheating.
Chapter 3 - System Setup
VENUE FIT™ SERVICE MANUAL
DAMAGE IN TRANSPORTATION
Recommended Procedure
on page x in
Chapter 1 -
Table 3-
3-11