Rear Casters Removal Procedure (cont'd)
5.) Activate the brakes on all Casters, except the one you intend to replace.
6.) Turn the Rear Caster that stands on the Bevel Edged Board in the direction as shown in
Figure
7.) The system is now nearly balanced between one Front and one Rear Caster (illustrated in this
Procedure; Left Front and Right Rear).
8.) Make the System rest on both Front Casters and lift the Rear Caster. Put the Wooden Wedge under
the AC Transformer. This will stabilize the System with the Rear Caster free from the floor.
Rear Wheel free
from the floor
Figure 8-89 System stabilized with Rear Caster free from the floor
11
8-88.
Figure 8-88 Activate Rear Brakes
Chapter 8 - Replacement Procedures
V
7 S
M
IVID
ERVICE
Wooden Wedge under
AC Transformer
ANUAL
8 - 93