Backup of Patient Archive and System Configurations to MO Disks (cont'd)
7.) Select Backup Now.
a.) The system will first check if the MO disk is inserted. If not, a dialogue window is displayed
b.) The system will then check if the MO disk is formatted. If not, a dialogue window is displayed
A progress dialogue is displayed showing the current operation being performed.
If the amount of data exceeds the capacity of the MO disk, the user is prompted to insert a new
(formatted) MO disk. At the end of the process, a status for each item is displayed in the Result column.
If errors occur during backup, a list of the patient records/ examinations that could not be backed up is
displayed.
8 - 42
Section 8-8 - Software Loading Procedure - Software Version v3.x
11
prompting you to insert an MO disk. Select OK to resume backup.
prompting you to format the disk.