Create a New Dataflow

3-9-16-1
Overview
It is possible to make new dataflows by combining the predefined settings. The table below describes
the legal combination of inputs and outputs in a dataflow.
Output
Input
Output Database
Database
Remote Database
DICOM Worklist
DICOM CD
DICOM MOD 5.25
DICOM MOD 3.5
Query/Retrieve
EchoPAC Mac
Worklist/database
Worklist/rem database
No input device
Table 3-16
3-9-16-2
Create a New Dataflow with Worklist, Database and DICOM Storage
(Introduced for software v4.0)
To create a new dataflow with Worklist, Database and DICOM Storage, do the following in exactly these
steps:
NOTICE
Failure to add destinations to the dataflow in the correct order may or will cause intermittent errors.
1.) Add the Worklist Input, highlight it, select properties, and name it (i.e.: DR PACS Worklist or just
Worklist).
Go to the IP Address field, scroll down and select <Modify>. This will allow you to edit or add IP
Addresses (New Destinations/devices). If the correct IP Address is already in the list, just select it.
2.) Add the Database Input/Output, highlight it, select properties, and name it (i.e.: Local Archive - Int
HD).
Go to the IP Address field and (if the Database is to be the Local Archive) verify that "My Computer"
is selected.
The IP Address should then be correct and the Database Name and Archive File Location entries
should auto-populate correctly.
11
DICOM
No
Remote
Database
X
X
X
X
X
X
X
Allowed combinations of inputs and outputs in a dataflow
Chapter 3 - System Setup
DICOM
MOD
MOD
DICOM
DICOM
5.25
3.5
Storage
Print
X
X
X
X
X
V
7 S
IVID
ERVICE
Remote
Database
Database
Database
+
+
+
DICOM
DICOM
DICOM
MOD
Storage
MOD
5.25/3.5
5.25/3.5
X
X
X
X
M
ANUAL
Remote
Database
+
DICOM
Storage
X
X
3 - 75