System Setup

Verify customer order

Physical inspection

System Voltage Settings
NOTE:
WARNING

EMI protection

3-22

Preparing for Setup

Compare items received by the customer to that which is listed
on the delivery order. Report any items that are missing, back
ordered, or damaged.
Verify that the system arrived intact (visual inspection).
If the system has been damaged, please refer to
Transportation' on page i-11
Verify that the scanner is set to the correct voltage. The
Voltage settings for the Venue 50 Scanner is found on a
label located on the AC adapter.
220-240VAC(China); 100-120VAC(USA/Japan);
220-240VAC(Europe, Latin America).
Check your local grid and confirm the voltage.
Connecting Docking Station/Cart to the wrong voltage
level will most likely destroy the system.
The Venue 50 has been designed to minimize the effects of
Electro-Magnetic Interference (EMI). Many of the covers,
shields, and screws are provided primarily to protect the system
from image artifacts caused by this interference. For this reason,
it is imperative that all covers and hardware are installed and
secured before the unit is put into operation.
See
'EMI limitations' on page 2-8
EMI protection.
in the beginning of this manual.
for more information about