GE Vivid 7 Service Manual page 221

Hide thumbs Also See for Vivid 7:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FC091194, R
IRECTION
EVISION
3-10-8-3
Verify the Connection to a Device - Software Version v3.x
1.) Select (highlight) the device you want to verify the connection to.
2.) Select CHECK to start the verification process of the connection to the device.
The verification process may use some amount of time (several seconds). When done, a sign in
front of the device indicates if the test passed (green check-mark) or failed (red X).
NOTE:
If the Check button fails directly, the AE Title is probably wrong.
If the Check button fails after a long time (corresponding to the timeout), the IP address or Port
Number is probably wrong.
11
Figure 3-105 Verify connection to a device
Figure 3-106 Verification result
Chapter 3 - System Setup
V
7 S
M
IVID
ERVICE
ANUAL
Select device to
be verified.
Check
A green check-mark
indicates that the test
passed.
A red cross (as
illustrated) indicates
that the connection
failed.
3 - 97

Advertisement

Table of Contents
loading