Label, Internal Connections (Int.conn.) - GE Vivid 7 Service Manual

Hide thumbs Also See for Vivid 7:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FC091194, R
IRECTION
EVISION
1-5-19

Label, Internal Connections (Int.Conn.)

Located on the outside of the Front-End Card Rack's cover (inside unit). Different versions of the label
have been used since production start.
Cable and Connector numbering and naming system used:
IIO
= Internal IO module
EIO
= External IO module
BEP
= Backend Processor
FEP
= Frontend Processor
ACP
= AC controller module
ACD
= AC distribution box
ACT
= AC isolation transformer module
BEPIO
= IO module on the BEP
Axx :
connectors on IIO with other connections than the BEP
Bxx :
connectors on IIO that are connected to BEP
Cxx :
connectors on BEP, except for the BEPIO
Dxx :
connectors on BEPIO
Exx :
connector on the BEP power s u p p l y
Fxx :
connectors on AC P
Gxx :
connectors on FEP
Hxx :
external accesable connectors (Doppler probe)
Kxx :
connectors on MODEM
Mxx :
connectors on ACT
Nxx :
connectors on ACD
Lxx :
connectors on EIO
F
connector
to EIO
A
25
P30
1
P31
A13
A12
A12
Spare power (future option)
A13
Power to 3D box (future option)
A14
AC power control
A15
Signals to rotation adapter (future option)
A16
Signals to 3D box (future option)
P32
A15
A14
A16
L4:
L5:
L6:
L7:
L8:
L9:
L10:
L11:
L12:
L13:
L14:
L15:
L16:
L17:
1 - 52
11
AC
Distribution
box
N5
N4
N3
N2
N1
H 1
G1
Audio out (right)
Audio out (lef t)
Trig ou t
External IO module
Foot swit ch
Serial port
L1
L2
Remote expose #1
Remote expose #2
Analog modem
Composite video output
B&W video outpu t
Service
Analogue
RS-232
S-Video output
use ONLY
phone
Connect
SVGA out p u t
(remote
Connect
USB
to InSite
Etherne t
to InSite
power on/off
MODEM
MODEM
control)
Figure 1-7 Label on FEP2 (w/RFI)
Section 1-5 - Labels Locations
Live Voltage inside,
AC power control modu l e
F1
F2
F3
F4
F5
115V
230V
230V
230V
115V
or
230V
115V or 230V inl e t
voltage selection
F6
From Isolation
transformer
230V
115V
To Isolation
transformer
Cable to
C15
G3
Isolation
transformer
G4
modul e
Board location in Front End Processor (FEP)
M
L9
L4
L5
L6
L8
L7
F
RESET
L11
L12
L13
L14
L15
L16
L3
External IO module
(Front v i e w)
V
7 S
IVID
WARNING !
do not open.
F9
230V
115V
Voltage Selec tion
on Peripherals
F8
F7
60,0
M1
L10
Modem
MT5634ZBA
Modem
power
K4
suppl y
L17
K3
K1
M
ERVICE
ANUAL
K2
Part no. : FC314311
Rev. : 01