Setting the DICOM Server as Default Dataflow - Software v2.x
1.) Finish the verification in
page
You need to have green check marks on both Hardware Ping and DICOM Ping before you continue.
1.) Select the DATAFLOW tab.
2.) Select "DICOM Server" dataflow on the "Dataflow Name" drop-down menu.
.
DATAFLOW tab
3.) Check the Default box.
.
4.) Exit Config and restart the scanner (by hitting CTRL/ALT/R at the same time).
11
"Check the Connection to the DICOM Server (Ping) - Software v2.x" on
3-130.
Figure 3-141 Selecting DICOM Dataflow Services
Figure 3-142 Set DICOM as Default Dataflow
Chapter 3 - System Setup
V
7 S
M
IVID
ERVICE
ANUAL
Dataflow Name
drop-down menu
Select
"DICOM Server"
Select this check box to
make DICOM the Default
dataflow