HL7 Communication Setup - Software v3.x

3-10-9-1
Introduction
The procedures below tell you how to set up the Vivid 7 so it can connect to the Hospital Information
System (HIS), via the Vivid HL7 Gateway. By connecting to the HIS, demographic information i.e Patient
ID, Name, Gender etc. can be pulled from the HIS, reducing "double-work" and typing errors.
Patient Reports can be sent back to the HIS for storage together with the rest of the Patient records.
3-10-9-2
Preparations
1.) Before you can set up the HL 7 Communication, you need to collect some information.
-
-
Typical source for this information is the Mitra support personnel.
Figure 2-2 "Pre-quote Worksheet for DICOM Network Information" on page 2-14
gathering this information.
2.) Provide the computer name of each EchoPAC PC Workstation and Vivid 7 being configured, to the
Mitra support personnel.
3-10-9-3
Configuration of Modality Worklist Services - Software v3.x
Identify Dataflow to be Used
1.) Identify which dataflow is to be used at the site. If they will be using DICOM Modality Worklist, then
the dataflow must include a Worklist service.
2.) On the Connectivity screen, select the Dataflow tab.
3.) Select the name of the site's dataflow from the pull-down.
4.) In the Services section, locate the Worklist service.
5.) Record the service's name and destination device.
Configure the DICOM Worklist Service
1.) Select the Services tab.
2.) Select the Worklist destination device from the pull down menu.
3.) Enter the IP address of the Vivid HL7 Gateway.
4.) Enter a descriptive name in the Name field. Example: Vivid HL7 Gateway. The use of this field is
optional).
5.) Enter AE Title. Use the Computer Name for the unit you are configuring.
6.) Enter Port Number. Port 3320 is default, but Mitra may have provided another port number.
7.) Set Time-out to 90.
8.) Leave the rest of the fields unchanged.
3 - 98
11
The IP address of the Vivid HL7 Gateway
Port Number used by the Vivid HL7 Gateway (Default port: 3320)