2D Mode (B mode) Checks
4-3-6-1
Introduction
The 2D Mode is the system's default mode.
4-3-6-2
Preparations
Use a phantom when doing these tests.
1.) Connect one of the probes, to the scanner's left-most active probe connector.
-
-
2.) Turn ON the scanner.
-
The 2D Mode window is displayed (default mode).
3.) If needed, adjust the Display's Brightness and Contrast setting.
4 - 32
11
Figure 4-27 2D Mode Screen Example
See
"Probe Connection" on page 3-21
See section
Section 9-26 "Probes" on page 9-80
See
4-2-2 "Power ON/ Boot UP" on page 4-3
Section 4-3 - Functional Checks
Status Window
Control Menu
for info about connecting the probes.
for available probes.