GE Vivid 7 Service Manual page 131

Hide thumbs Also See for Vivid 7:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FC091194, R
IRECTION
EVISION
3-3-4
Receiving the Vivid 7
3-3-4-1
Examine All Packages
Examine all packages closely at time of delivery, as described in the procedure below.
Table 3-5
STEP
1
2
3
4
11
Examine All Packages
TASK
Is damage apparent?
If yes; continue with the instructions in subsection
3-3-4-2 - Damage in
Transportation.
If no; continue with the next step.
Is the Shock Indicator red colored inside the middle of the
indicator?
If yes: The Shock Indicator has been activated.
Continue with the instructions in subsection
3-3-4-2 - Damage in
Transportation.
If no: Continue with step 2.
Is the Tilt Indicator red colored inside the middle of the
indicator?
If yes: The Tilt Indicator has been activated.
Continue with the instructions in subsection
3-3-4-2 - Damage in Transportation
with the next
step.
If no: Continue with the instructions in
Vivid
7.
The two upper hinges on the Front Side and the Rear Side of
the wooden transportation bow have been sealed with red
plastic seals, marked GE Vingmed Ultrasound and a serial
number. Verify that the four red plastic seals are intact at
arrival.
If seals are broken:
If broken, it indicates that the container may have been
opened after it left the manufacturer.
Continue with the instructions in subsection
3-3-4-2 - Damage in
Transportation.
Continue with the instructions in
Chapter 3 - System Setup
ILLUSTRATIONS
RED COLOR
RED COLOR
before you continue
3-3-5 - Unpacking
3-3-5 - Unpacking Vivid
7.
V
7 S
M
IVID
ERVICE
ANUAL
3 - 7