Getting Started
CAU T ION
CAU T ION
CAU T ION
CAU T ION
CAU T ION
14
VENTILATION REQUIREMENTS: When installing the product in a
cabinet, the convection into and out of the product must not be
restricted. The ambient temperature (outside the cabinet) must be
less than the maximum operating temperature of the product by 4 °C
for every 100 watts dissipated in the cabinet. If the total power
dissipated in the cabinet is greater than 800 watts, then forced
convection must be used.
Install the instrument so that the detachable power cord is readily
identifiable and is easily reached by the operator. The detachable
power cord is the instrument disconnecting device. It disconnects
the mains circuit from the mains supply before other parts of the
instrument. The front panel switch is only a standby switch and is
not a LINE switch. Alternatively, an externally installed switch or
circuit breaker (which is readily identifiable and is easily reached by
the operator) may be used as a disconnecting device.
Always use the three-prong AC power cord supplied with this
instrument. Failure to ensure adequate earth grounding by not using
this cord may cause instrument damage.
This instrument has autoranging line voltage input. Be sure the
supply voltage is within the specified range.
Use of controls or adjustment or performance of procedures other
than those specified herein may result in hazardous radiation
exposure.
General Safety Considerations
Agilent 86038A Optical Dispersion Analyzer, Third Edition