Troubleshooting GPIB Communications Problems

1. Check the ODA Status Log
2. Ensure that GPIB cables are configured correctly.
3. Check the system GPIB Addresses.
Table 11
Agilent 86038A Optical Dispersion Analyzer, Third Edition
Troubleshooting GPIB
Communications Problems
If the ODA application is running, check the ODA Status Log for any
warnings or errors. Press
Check the GPIB cables, they should match
The system is configured to work with set GPIB addresses for the tunable
laser source and multi-wavelength meter (when applicable). If the GPIB
addresses for these instruments have been changed, the ODA application
will not be able to communicate with them.
The default address for an 8164B and an 86122A is the same — 20. If both
instruments are on the bus with the same GPIB address, you may get
confusing error messages, likely a message that says the ODA Application
cannot contact the tunable laser source.
Follow the details below to check and reset the GPIB addresses on the
tunable laser source and multi-wavelength meter. The GPIB address
cannot be changed from the ODA software, but must be altered on the
individual instruments.
Instrument
Lightwave Measurement System
with Tunable Laser Source Module
Multi-Wavelength Meter
to view a list of status messages.
Status Log
Figure 2 on page
Model #
GPIB Address
8164B /
20
81640B or
81600B
86122A
10
Troubleshooting and Upgrades
19.
Path to set GPIB Address
Config > GPIB Address
Setup > GPIB Setup