Cleaning Connections for Accurate Measurements
Cleaning Connectors
CAU T ION
Table 13 Cleaning Accessories
Table 14 Dust Caps Provided with Lightwave Instruments
CAU T ION
Agilent 86038A Optical Dispersion Analyzer, Third Edition
The procedures in this section provide the proper steps for cleaning fiber-
optic cables and Agilent Technologies universal adapters. The initial
cleaning, using the alcohol as a solvent, gently removes any grit and oil. If
a caked-on layer of material is still present, (this can happen if the
beryllium-copper sides of the ferrule retainer get scraped and deposited on
the end of the fiber during insertion of the cable), a second cleaning should
be performed. It is not uncommon for a cable or connector to require more
than one cleaning.
Agilent Technologies strongly recommends that index matching
compounds not be applied to their instruments and accessories.
Some compounds, such as gels, may be difficult to remove and can
contain damaging particulates. If you think the use of such
compounds is necessary, refer to the compound manufacturer for
information on application and cleaning procedures.
Item
Cotton swabs
Small foam swabs
Item
Laser shutter cap
FC/PC dust cap
Biconic dust cap
ST dust cover
To clean a non-lensed connector
Do not use any type of foam swab to clean optical fiber ends. Foam
swabs can leave filmy deposits on fiber ends that can degrade
performance.
Troubleshooting and Upgrades
Agilent Part Number
8520-0023
9300-1223
Agilent Part Number
08145-64521
08154-44102
08154-44105
1401-0291