Appendix
General data
Temperature measurement
Installation
Supplementary documentation
80
CLS 52 sensor data
Lower measuring range limit (non-compensated)
Reaction time (T
)
90
Temperature sensor
Temperature response time
Heat transfer bushing with Pt 100
Material
Seal
Cell constant k (measurement deviation) ±0.5 %)
Storage temperature
Protection type (DIN 40050)
Measurement deviation (–5 ... +100 °C)
Measurement deviation (+100 ... +140 °C)
Medium temperature
Ambient temperature
for sterilisation
Pressure
Measuring sensor material
Surface roughness
Required pipe cross-section
Dairy pipe fitting, clamp nozzle G 1½
APV, Varivent connection
Technical Information CLS 52
Mycom CLM 152
0 µS/cm ... 2000 mS/cm
< 3 seconds over total measuring range (single-circuit
instrument)
Pt 100 Class A as per IEC 751
t
... < 15s
90
V4A (1.4571)
O-ring, EPDM
-1
5.9 cm
–25 ... +80 °C
IP 67
± (10 µS/cm + 0.5 % MR)
± (30 µS/cm + 0.5 % MR)
–5 ... +125 °C
–10 ... +70 °C
+140 °C (max. 30 min)
max. 16 bar (90 °C)
PEEK
Ra ≤ 0.5 µm
³ DN 65
≥ DN 40
order no. 50086110
Specifications subject to change
Endress+Hauser