Mycom CLM 152
Conductive sensors
Caution:
Make absolutely sure you protect
plugs and terminals from moisture.
Moisture leads to inaccurate
measuring results!
Special measuring cable required for connecting conductivity sensors
Sensor type
2-electrode sensor with or without
temperature sensor Pt100
Inductive sensor CLS 50
Conductivity measurement conductive max. 100 m with CYK 71 (equivalent to 10 nF)
Resistance measurement
11 12 13
gn
gn wt ye
3
ϑ
CLS 12 / 13
11 12 13
gn wt
gn wt ye
gn
bn
ϑ
CLS 30
Endress+Hauser
Cable
CYK 71 (0.04 Ω/m)
Fixed cable on sensor
Maximum cable length
max. 20 m with CYK 71 (equivalent to 2 nF)
14
15
16
17
wt
ye
bk
CYK 71
bk
4
2
1
CLS 15 / 19 / 20 / 21
14
15
16
17
ye
bk
CYK 71
bk
wt
ye
CLS 50 / 52
Note:
•
The instrument has a function to
offset cable length and capacitance
to compensate for cable resistance
(see Chapter 11.5.5 „Determine
cable resistance").
•
A waiting time must be observed
depending on the sensor used
and the medium temperature
(thermal matching) until the sensor
supplies the correct temperature
readings.
Extension
VBM box + CYK 71
VBM box + CLK 5
11 12 13
14 15 16 17
gn wt ye
bk
CYK 71
gn wt
ye
bk
3
1
2
ϑ
11 12 13
14 15 16 17
gn wt ye
bl rd
wt
rd
CLK 5 or fixed cable
ϑ
Installation
lm152e04.chp
rd bl
Connection of conductive
Fig. 4.26
and inductive sensors
AN-CLS.CDR
23
Need help?
Do you have a question about the Mycom CLM 152 and is the answer not in the manual?
Questions and answers