Endress+Hauser Mycom CLM 152 Operating Instructions Manual page 42

Conductivity measuring transmitter
Hide thumbs Also See for Mycom CLM 152:
Table of Contents

Advertisement

Instrument configuration
40
7.1.5
General settings
Function
Select language
1)
Input damping
(size of time window over which
measurement is averaged;
2 measured values per second)
Temperature unit
Set date
Set time
Tag number (max. 32 chars.)
(measuring point designation)
Display contrast (E+H logo)
1)
Input damping:
To increase immunity of measurement to interference, a continuous input damping function can
be switched on in the form of an input filter (averaging function). The filter length must be
adapted by trial and error to the process so that, although transients are suppressed, any
actual changes in the measured value are recorded.
Selection
Language version A:
Deutsch, English, Français, Italiano
Language version C:
Deutsch, English, Français,
Nederlands, Japanese
Filter length, 0 (= off) ... 30
Celsius [°C]
Fahrenheit [°F]
Kelvin [K]
Weekday, day, month, year
Hour, minute
0 ... 9; A ... Z
Set for optimum view
Mycom CLM 152
Factory setting
English
0
Celsius [°C]
Endress+Hauser

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Mycom CLM 152 and is the answer not in the manual?

Questions and answers

Table of Contents