Instrument diagnostics
70
11.3
Error table
No.
E001
Fault in internal communication
E002
Data error in EEPROM
E003
Invalid configuration
E004
Changed configuration
E005
Unknown card code
E006
Checksum error in EEPROM
Transmitter faulty or transmitter function
E007
disturbed
E010
Temperature sensor defective
E011
Temperature sensor 2 defective
E015
Current loop 1 open
E016
Current loop 2 open
Sensor error (inductive only)
E017
E018
Sensor error channel 2 (inductive only)
E036
Calibrating range of sensor 1 exceeded
E037
Calibrating range of sensor 1 undershot
E038
Calibrating range of sensor 2 exceeded
E039
Calibrating range of sensor 2 undershot
E040
TC table invalid
E041
Substance table invalid
Adaptation factor range of sensor 1
E049
exceeded
Adaptation factor range of sensor 1
E050
undershot
Adaptation factor range of sensor 2
E051
exceeded
Adaptation factor range of sensor 2
E052
undershot
Failure
Display
Send instrument to Endress+Hauser
agency responsible for repair or call
Service.
If error »07« occurs at the same time,
check transmitter acc. to this error.
Check configuration of slots
Accept new configuration using "Set config"
in "Service / Factory settings" menu
Check configuration of slots
Carry out checksum correction in "Special
functions"
Check temperature measurement and
terminals; poss. check measuring
transmitter and measuring cable with
temperature simulator
Check terminals, lines and poss.
connected devices
Send sensor to your Endress+Hauser
agency for examinaton or request Service
Maintenance required
Recalibrate sensor; poss. check sensor
and terminals; check measuring
transmitter and measuring cable with CD
simulator
Correct temperature compensation table;
check temperature measurement and
terminals; poss. check measuring
transmitter and measuring cable with
temperature simulator
correct substance table
Enlarge the distance of sensor from wall
Enlarge the distance of sensor from wall
Enlarge the distance of sensor from wall
Enlarge the distance of sensor from wall
Mycom CLM 152
Corrective action
Check sensor connection
• Check if measuring range is exceeded
• Check on exceeded sensor cross-talk
• Check temperature measurement. If
temperature measurement is not in
order, replace the module