Endress+Hauser Mycom CLM 152 Operating Instructions Manual page 59

Conductivity measuring transmitter
Hide thumbs Also See for Mycom CLM 152:
Table of Contents

Advertisement

Mycom CLM 152
The USP function can be activated from
the Instrument data/USP menu.
After activation, you can make the following
settings:
Function
On delay
Hysteresis
1)
Cell
MR 1
k = 0.01
0.0 ... 200.0 nS/cm
k = 0.1
0.000 ... 2.000 µS/cm
k = 1
0.00 ... 20.00 µS/cm
k = 10
0.0 ... 200.0 µS/cm
1)
The selection of measuring range is executed automatically with internal measuring steps to ensure optimum accuracy
and resolution.
Endress+Hauser
Selection
0 ... 60 s
Setting range sensor-dependent
(see Table)
1)
MR 2
MR 3
0.200 ... 2.000 µS/cm
2.00 ... 20.00 µS/cm
2.00 ... 20.00 µS/cm
20.0 ... 200.0 µS/cm
20.0 ... 200.0 µS/cm
200 ... 2000 µS/cm
200 ... 2000.0 µS/cm
2.00 ... 20.00 mS/cm
Limit configuration
Factory setting
0 s
0 µS/cm
1)
1)
MR 4
20.0 ... 200.0 µS/cm
200 ... 2000 µS/cm
2.00 ... 20.00 mS/cm
20.0 ... 200.0 mS/cm
lm152e08.chp
57

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Mycom CLM 152 and is the answer not in the manual?

Questions and answers

Table of Contents