Instrument configuration
46
Example of a current output table for log 3
Current output [%]
Conductivity [µS/cm]
0
5
10
15
20
25
30
35
40
45
50
Programming the current output table:
Function
Select support point
Enter value for current output
(%) and conductivity
Status display
Current output [%]
0.1
55
0.14
60
0.2
65
0.28
70
0.4
75
0.56
80
0.79
85
1.12
90
1.58
95
2.24
100
3.16
Selection
Enter number of support point
2 ... 21
0 ... 100 %
0 ... Measuring value final
value (dependent on
connected sensor, see p. 43)
Current output and
conductivity value must
increase from element to
element, otherwise an error
message will appear:
• „Percentage values
monotone non-increasing»
or
• „CD values monotone
non-increasing
Mycom CLM 152
Conductivity [µS/cm]
4.47
6.31
8.91
12.6
17.8
25.1
35.5
50.1
70.8
100
Factory setting