Calibration Presettings; Clean Function - Endress+Hauser Mycom CLM 152 Operating Instructions Manual

Conductivity measuring transmitter
Hide thumbs Also See for Mycom CLM 152:
Table of Contents

Advertisement

Instrument configuration
50
7.4

Calibration presettings

→ Instrument data
PARAM
→ Calibration
Function
Hold for calibration
Enter temperature coefficient Tc of
calibration solution
Enter temperature of calibration
solution (only for MTC)
7.5

Clean function

→ Instrument data
PARAM
The »Chemoclean« clean function requires a
device with a total of five output contacts.
The CLA 111 assembly and the CLR 30 spray
head are required to automate the cleaning
functions.
Contact 3 (terminal 91/92) is used for the
motive water, contact 4 (terminal 93/94) is for
dosing of cleaner.
Selection
yes, no
0.00 ... 10.00 %/K
–35.0 ... +250.0 °C
→ Chemoclean
The »Chemoclean« clean function must be
set in the »System configuration / Output
contacts« menu (see Chapter 7.1.4).
Mycom CLM 152
Factory setting
yes
2.10 %/K
25.0 °C
Endress+Hauser