Instrument diagnostics
72
No.
Display
E080
Range for current output 1 too small
E081
Range for current output 2 too small
E142
Knee point outside current output range 1
E143
Knee point outside current output range 2
Current output range 1 too small for
E144
selected MR (= measuring range)
Current output range 2 too small for
E145
selected MR (= measuring range)
Knee point outside current output
E148
range 1 (MBX = current measuring range) Correct configuration in "Current outputs"
Knee point outside current output
E149
range 2 (MBX = current measuring range)
E150
Measured value outside USP table
E151
Temperature outside limits for USP table
Warnings
Corrective action
Enlarge range in "Current outputs"
Correct cnfiguration in "Current outputs"
Make range in "Current outputs" menu
larger
menu
Mycom CLM 152
Endress+Hauser
Need help?
Do you have a question about the Mycom CLM 152 and is the answer not in the manual?
Questions and answers