Endress+Hauser Mycom CLM 152 Operating Instructions Manual page 63

Conductivity measuring transmitter
Hide thumbs Also See for Mycom CLM 152:
Table of Contents

Advertisement

Mycom CLM 152
Determining the adaptation factor
Function
Select sensor (not for single-circuit)
Message
Display conductivity of calibration
solution
current Tc, temperature
Enter setpoint of calibration solution
Display adaptation factor
Calibrate
Return to Measuring mode
Calibrate sensor 1
Return to Select sensor (above)
If »Cal. channel 2« selected (only for difference)
Calibrate sensor 2
Return to Select sensor (above)
Note:
You can lead the air set in menu
service / special function /
optimisation
Endress+Hauser
Selection
Sensor 1
Sensor 2
0 µS/cm ... 1000 mS/cm
End Cal.
Repeat Cal.
Cal. channel 2 (only for difference)
If »End Cal.« selected
If »Repeat Cal.« selected
Calibration
Default
Sensor 1
Leave sensor in
process
Current
measured value
xx.xxx
End Cal.
Sensor 1
Sensor 2
lm152e09.chp
61