Limit configuration
Selection
µ
CLS 50
0
S/cm 1000 mS/cm
µ
CLS 52
0
S/cm 1000 mS/cm
k = 0.01
0 nS/cm 200.0
µ
k = 0.1
0
S/cm 2000
µ
k = 1
0
S/cm 20.00 mS/cm
54
8.1
Instruments with two contacts
→ Instrument data
PARAM
→ Limit contactor
Function
Select group
Switch on/off output
Limit
Hysteresis
Contact function
On delay
Off delay
Contact type
Alarm threshold
Alarm delay
Switch over operating mode
Limit contactor manual mode
(if „Manual" selected)
Cell dependency
Limit
Default
Limit 1: 50.0 mS/cm
Limit 2: 950 mS/cm
Limit 1: 50.0 mS/cm
Limit 2: 950 mS/cm
µ
Limit 1: 10.00
S/cm
µ
S/cm
µ
Limit 2: 190.0
S/cm
µ
Limit 1: 100.0
S/cm
µ
S/cm
µ
Limit 2: 1900
S/cm
µ
Limit 1: 1000
S/cm
Limit 2: 19.00 mS/cm
Limit configuration
Alarm configuration
Operating mode
If „Limit configuration" selected
on, off
see Table "Cell dependence"
see Table "Cell dependence"
Min function
Max function
0 ... 7200 s
0 ... 7200 s
NC contact, NO contact
If „Alarm configuration" selected
see Table "Cell dependence"
0 ... 6000 s
If „Operating mode" selected
Auto limit contactor 1
Manual limit contactor 1
off, on
Hysteresis
Selection
Default
µ
0.1
S/cm 200.0 mS/cm
20.00 mS/cm
µ
1
S/cm 200.0 mS/cm
20.00 mS/cm
µ
0.1 nS/cm 20.00
S/cm
2.000
µ
µ
0.001
S/cm 200.0
S/cm
20.00
µ
µ
0.01
S/cm 2000
S/cm
200.0
Mycom CLM 152
Selection
Alarm
Selection
µ
0.1
S/cm 1000 mS/cm
µ
1
S/cm 1000 mS/cm
µ
µ
S/cm
0.1 nS/cm 200.0
µ
µ
µ
S/cm
0.001
S/cm 2000
µ
µ
S/cm
0.01
S/cm 20.00 mS/cm
Factory setting
Limit
configuration
off
Min function
0 s
0 s
NO contact
0 s
Manual
limit contactor 1
off
Default
50.0 mS/cm
50.0 mS/cm
µ
S/cm
10.0
S/cm
µ
S/cm
100.0
S/cm
µ
1000
S/cm
Endress+Hauser