Acquity Upc 2 Photodiode Array Detector Specifications; Sample Manager - Fixed Loop Specifications - Waters ACQUITY UPC2 System Manual

Hide thumbs Also See for ACQUITY UPC2:
Table of Contents

Advertisement

B.8
ACQUITY UPC
specifications
Table B–15:
Item
Data rate
Digital filter
Digital resolution
a
Drift
(dry)
Linearity
Nitrogen purge
Noise (wet)
Optical resolution
Order filter
Photodetector
Wavelength accuracy
Wavelength calibration filter
Wavelength range
Wavelength repeatability
a. ASTM drift tests require a temperature change of less than 2 °C/hour (3.6 °F/hour) over a one hour period. Larger
ambient changes will result in larger drift. Better drift performance depends on better control of temperature fluctuations.
To realize the highest performance, minimize the frequency and the amplitude of temperature changes to below
1 °C/hour (1.8 °F/hour). Turbulences around one minute or less can be ignored.
B.9

Sample manager - fixed loop specifications

Table B–16:
Feature
Integrated leak management
Operating flow rate range
Maximum operating pressure
pH range
Unattended operation
2
Photodiode array detector
PDA performance specifications
Specification
≤ 80 Hz, maximum
Variable with data rate
1.2, 2.4, 3.6, 4.8, 6.0, 7.2, 8.4, 9.6, 10.8, 12.0 nm
≤1000 µAU/hr (at 230 nm, 2 Hz, 1 sec TC, 3.6 nm res, dry
analytical flow cell); 2 hr warm-up, environmental stability <±2
°C/hr variation
≤5% at 2.0 AU, propylparaben series at 257 nm, 10 mm
analytical flow cell
Purge fitting present on optics bench and rear panel
<10 µAU (254 nm, 2 Hz, 1 sec TC, 3.6 BW res, 10 mm analytical
flow cell) flow rate 0.5 mL/min, 90:10 H
1.2 nm (nominal), with 50-µm slit
Fixed, 370 nm to 800 nm
512 element selective access photodiode array
±1.0 nm (tested with Erbium Perchlorate)
Erbium filter, used at startup or on-demand
190 to 800 nm
±0.1 nm
SM-FL performance specifications
Specification
Leak sensors, as standard, and safe leak handling. Drip
trays direct all leaks to the front of the instrument, and then
into the waste line.
0.010 to 4.000 mL/min, in 0.001 mL increments.
41,369 kPa (413 bar, 6000 psi) up to 3.25 mL/min.
pH 2 to 12.
Leak sensors, full 96-hour diagnostic data display through
the console software.
November 16, 2016, 715004521 Rev. C
Page 101
O:acetonitrile
2