Attach the epi-fluorescence cube turret and epi-fluorescence attachment (required for
14
epi-fluorescence microscopy).
Tool: Hex driver (2 mm across flats)
Hex wrench (3 mm across flats)
First, connect the epi-fluorescence cube turret and
epi-fluorescence attachment and then attach them on the
microscope arm.
(1)
Insert the positioning pin of the epi-fluorescence
attachment while aligning it with the positioning hole
on the epi-fluorescence cube turret side, and then
tighten the connection fixing screws (x2) on the
epi-fluorescence cube turret using a hex driver.
(2)
For motorized and intelligent fluorescence cube
turret, remove the contact cover of the contact arm.
(When the manual epi-fluorescence cube turret is
used, do not remove the contact cover.)
(3)
Loosen the fixing screw on the front of the arm using
a hex driver so that the tip of the screw does not
protrude into the connecting section.
(4)
Align the round dovetail and convex contact on the
bottom of the epi-fluorescence cube turret with the
round dovetail and concave contact of the arm and
slide the entire mounted attachment to the rear.
(Manual fluorescence cube turret does not have a
convex contact.)
(5)
Tighten the fixing screw loosened in step (3).
(6)
Tighten the four fixing screws on top of the
epi-fluorescence illuminator using a hex wrench.
(7)
Attach the HG adapter to the bayonet mount at the
rear of the epi-fluorescence attachment and connect
the HG precentered fiber illuminator. (See the
instruction manual provided with the HG
precentered fiber illuminator.)
Chapter 1
Assembly
Connection fixing
screw
Connecting epi-fluorescence cube turret to
Fixing screw
Attaching and fixing epi-fluorescence cube turret
F. ST
OP
21
Positioning
Epi-fluorescence
hole
cube turret
F. ST
OP
A. ST
OP
epi-fluorescence attachment
P
F. STO
2
3
4
5
6
FN- S2N
JAPA
N
and epi-fluorescence attachment set
Bayonet mount
HG adapter
A. ST
OP
Attaching HG adapter
Epi-fluorescence
attachment
P
A. STO
Epi-fluorescence
attachment fixing
screw
Contact
cover
HG fiber