Service and troubleshooting
March 2021
Table 5-5: Troubleshooting Chart for F
(continued)
Symptom
Wrong unit appears
on the display
Cannot commission
the display to
F
OUNDATION
segment
5.3.4
Device errors
Table 5-6
about the different error types can be found in Input registers 1100 - 1134 as shown in
Table
Table 5-6: Device Errors
Message
RAM Error
FPROM Error
108
Possible cause
Incorrect
configuration in AMS
Device Manager.
Measurement unit
that was chosen in
the Display Setup
window does not
match selected unit
in the Manual Setup
window
Missing Device
Description (DD)
Fieldbus
shows a list of error messages for the Rosemount 2230. Detailed information
5-6.
Description
Input register no. 1100
The following bits indicate a
serious RAM problem.
Bit 0: RAM
Input register no. 1102.
The following bits indicate a
serious FPROM problem or
wrong software versions
loaded.
Bit 0: Checksum Error
Bit 4: Boot Checksum
Bit 5: Boot Version (Invalid
version number)
Bit 6: Application Checksum
Bit 7: Application Version
(Invalid version number)
Fieldbus Related Problems
OUNDATION
Action
Ensure that the "Default" option is selected for
Unit in the Manual Setup window
Add Rosemount 2230 DD to the FF host
Action
(1)
.
Contact Emerson Automation Solutions /
Rosemount TankGauging service
department.
Reference Manual