Product certifications
March 2021
B.4

North America

B.4.1
I5 USA Intrinsic Safety
Certificate
Standards
Markings
Entity parameters
FISCO parameters
Specific Conditions for Safe Use (X):
1. The non-metallic parts incorporated in the enclosure of this equipment may
2. The enclosure contains aluminum and is considered to present a potential risk of
3. The Rosemount 2230 Graphical Field Display will not pass the 500Vrms dielectric
4. To maintain ingress protection ratings IP66 and IP67, PTFE tape or pipe dope is
148
FM17US0035X
FM Class 3600 – 2011
FM Class 3610 – 2010
FM Class 3810 – 2005
ANSI/NEMA 250 – 2008
ANSI/IEC 60529 – 2004
ANSI/ISA 61010-1:2004
ANSI/ISA 60079-0 – 2013
ANSI/ISA 60079-11 – 2013
IS/I,II,III/1/ABCDEFG/T4 Ta = -50 °C to +70 °C
Control Dwg D9240040-949
I/0/AEx ia IIC Ga T4 Ta = -50 °C to +70 °C
Control Dwg D9240040-949
Type 4X; IP66, IP67
Ui
30 V
17.5V
generate an ignition-capable level of electrostatic charge. Therefore particularly
when it is used for applications that specifically require Division 1 and Group II, Zone
0 located equipment, the equipment shall not be installed in a location where the
external conditions are conducive to the build-up of electrostatic charge on such
surfaces. Additionally, the equipment shall only be cleaned with a damp cloth.
ignition by impact or friction. Care must be taken during installation and use to
prevent impact or friction.
strength test and this must be taken into account during installation.
required for cable entries and blanking plugs.
Ii
Pi
300 mA
1.3 W
380 mA
5.32 W
Reference Manual
00809-0100-2230
Ci
Li
2.1 nF
1.1 µH
N/A
N/A
Reference Manual