Holding Registers store various device parameters used to control measurement
performance.
Procedure
1. Start the TankMaster WinSetup program.
2. In the TankMaster WinSetup workspace window, select the device icon.
3. Right-click and select View Input/View Holding Registers option, or from the
4. In the Registers Type list, select Predefined or All.
5. For the All option, you have to specify a range of registers by setting a start value in
6. The Registers Scope drop-down list has three options:
7. In the Show Values in pane, choose the appropriate register format Decimal or
8. Click the Read button.
Related information
Viewing input/holding registers using AMS Device Manager
Rosemount 2230 Graphical Field Display
Service menu choose Devices → View Input/View Holding Registers.
Now the View Input/Holding Register window appears.
Option
Predefined
All
the Start Register input field, and the total number of registers to be displayed in
the Number of Registers field (1-500). Up to 50 registers is recommended for a
quick update of the list.
Scope
Description
Basic
Standard setting that includes the most
commonly used registers
Service
Includes a wider range of registers for
advanced service and troubleshooting
Developer
For advanced users only
Hexadecimal.
Now the View Input/Holding Registers window is updated with the current register
values.
Description
View a basic selection of registers.
View a range of registers by your own choice (for
advanced service).
Service and troubleshooting
March 2021
Access level
View Only
Supervisor
Administrator
99