5.7.3
Viewing device status in AMS Device Manager
To view the current device status:
Procedure
1. Start AMS Device Manager and open the View → Device Connection View.
2. Double-click the FF network icon and expand the network node to view the devices.
3. Right-click or double-click the desired device icon to open the list of menu options.
4. Choose Service Tools.
5. In the Navigation Pane, select the Maintenance option.
6. Select the Details tab and click the Device Status button.
122
In the Device Status tab, the current status of the Rosemount 2230 is grouped in
separate categories. In the General pane check boxes indicate the current status of
the Rosemount 2230.