Configure Verify

Define the verify gas value and the acceptable range of readings for a verify.
If during a verify opera-
tion (Initiate Verify menu
option) the system finds
that the verify gas is not
within the acceptable range, a
Verify Failure message will
appear on the bottom line of
the display, and the Service
Alarm will trip. This mes-
sage will be cleared when a
successful verify is run (the
verify gas is within the range
specified), or a calibration is
run.
5-10 | Dycor CG1000-RTP Oxygen Analyzer
Entering Verify Gas Values
1. From the Calibrate Key menu, select Configure Verify.
2. Choose whether you want to enter the span gas value using a percent
or PPM scale.
3. Enter the verify gas value, then press ENTER. If you want to keep the
current zero gas value unchanged, press ENTER without entering a
value.
4. Enter the acceptable verify range. For example, if you enter a verify
gas range of 1%, and the verify gas value is 10%, then the acceptable
range is between 11% and 9%.
Verify Gas = xx
New Value?