Ametek Dycor CG1000-RTP User Manual page 19

Table of Contents

Advertisement

Sensor
operating Range
accuracy
Response time
Repeatability
Leak Rate
ambient
temperature
Humidity
Max. inlet
temperature
Sample flow
environment
power
Requirements
inlet pressure
calibration Gas
Requirements
Display
Recorder output
SpecificationS
0.1 PPM O
to 100% O
2
2
Percent: ±2% of reading or .05% O
PPM: ± 2% of reading or .5 PPM O
< 5 seconds at 150 sccm over one decade.
Percent: ± .5% of reading or .1% O
PPM: ± .5% of reading or .1 PPM O
3 x 10
atmosphere cc/sec
-8
5° C to 40 °C (41 °F to 104 °F)
10 - 80%, non-condensing
160 °F (70 °C)
125 to 200 sccm (150 sccm is recommended).
IEC Installation Category II
IEC Pollution Degree 2
Altitude: 2000 meters
115 VAC, 60 Hz, 150 VA
230 VAC, 50 Hz, 150 VA
Limit: 3 PSIG
Use calibration gases @ 150 sccm (limit is 125 to 200 sccm)
Zero Gas: From 0.1 PPM to 10% O
Span Gas: Minimum one decade above zero gas (10 times greater
recommended)
4 line x 20 character vacuum fluorescent. Displays combination of
oxygen (0.1 ppm O
to 100%, autoranging), time and date, cell tem-
2
perature, user-programmable text, thermocouple mV, cell mV, flow,
and pressure. Password protection and context-sensitive help are
also provided.
Two isolated linear current outputs. Select O
mocouple mV or cell mV. each output can be 4-20 mA, 0-20 mA, 20-4
mA or 20-0 mA, and is fully scalable. Hold or track during calibration
and select degree of damping. Maximum load is 1200 Ohms.
absolute, whichever is greater.
2
absolute, whichever is greater.
2
absolute, whichever is greater.
2
absolute, whichever is greater.
2
, balance N
2
2
, cell temperature, ther-
2
Specifications
|
2-1