RS-232 Communications Connections

The CG1000-RTP communicates with other computers using RS-232 serial
communications protocol. The connections used for RS-232 communica-
tions are on connector J7 and labeled as follows (see Figure 3-1 for location
of J7 on back of CG1000-RTP).

Select 115/230 Volt Operation

To change the selected voltage:
• Disconnect the AC line power cord from the analyzer.
• Open the top cover of the power entry module using a small-blade
screwdriver.
• Select the voltage by rotating the fuse holder so the desired voltage
aligns with the window, then snap into place.
If you later want to switch the voltage source for the analyzer, you will
also need to change the fuses in the power entry module:
115 VAC - Fuse Rating: 2 amp, 250 VAC - Slow Blow
230 VAC - Fuse Rating: 1 amp, 250 VAC - Slow Blow
The fuse sizes are 1/4" x 1 1/4". Do not use 5 x 20 mm fuses.
If the analyzer is a 230-Volt type, the power cord plug must be
changed to the proper type to match your country and/or electrical
code.
NOTE
dcd
Pin 1
Rd
Pin 2
Td
Pin 3
dTR
Pin 4
gnd
Pin 5
dsR
Pin 6
RTs
Pin 7
cTs
Pin 8
n/A
Pin 9
|
Installation & Start-Up