System Start-Up

Plug the power cord into the power entry module on the back of the
1.
analyzer.
Plug the other end of the power cord into the appropriate Ac power
2.
outlet.
Turn on the analyzer by pressing on the side of the power switch
3.
marked 1. The power switch is located on the back of the analyzer.
Adjust the flow of the sample gas to the desired level (see Chapter 2 -
4.
Installation for flow requirements).
5.
Allow the analyzer to warm-up for at least one hour.
6.
set up your alarms, analog outputs, and serial communications.
7.
Perform a regular calibration.
8.
Read the sample gas.

Password Restrictions

If there is a password restriction, you must enter the password before you
can access menu functions that can change system settings.
Display
You can define which information to place on the first three lines of the
display using the Display option from the Setup key. You can also use the
Display menu option to place your own text message on one of the first
three lines.
The last display line is reserved for system and error messages.
NOTE
If an alarm condition exists, alarm information in the form of an Up
Down
arrow (corresponding to a high or low alarm) will appear next to
the process reading on that display line.

Auto Ranging Feature

The CG1000-RTP analyzer includes an auto ranging feature. If the oxygen
reading falls below 0.5% O
wise, the system always displays readings using the percent scale.
, the display switches to the ppm scale. Other-
2
|
Installation & Start-Up
or
3-11