A thermal calibration is not a
regular calibration.

System Serial #

View your system serial # and manufacturing #. Service personnel may ask for
this information.
1. From the Setup menu, select System Serial # .
2. Select System Serial # or Manufacturing #.
3. The system displays the system serial # or manufacturing #.

Primary Calibration

Perform a primary calibration on the zirconium oxide oxygen sensor.
Always attempt to perform a regular calibration before resorting to a
primary calibration for all cases other than when you install a new cell.
A primary calibration should only be done in the following cases:
when you replace your zirconium oxide oxygen cell;
when your system fails a regular calibration due to a deteriorating cell.
Regular calibration isn't working
• Check to see if your calibration gas values match the values written on
the gas cylinders.
• Ensure that the correct flow of gases has been set and that the gases
are allotted enough time to flow through the sensor.
If these checks have been done and the system still can't be properly cali-
brated you should do a primary calibration.
Performing a primary calibration
Before starting the primary calibration, make sure you have properly
entered your calibration gas values using the Cal Gas Values option
NOTE
from the Calibrate key.
1. From the Setup key, select Primary Calib.
2. At this point, the system works as it does for a regular calibration.